Product Wiki · Recalls · U.S. FDA — Device Enforcement

IDM-MICRO-R. Electrical wheelchair component. — recall Z-2456-2025

U.S. FDA — Device Enforcement · Belgium · 2025-07-10 · Class I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2456-2025
Date of alert
2025-07-10
Country / jurisdiction
Belgium
Risk level
Class I
Notice type
recall
Brand
mo-Vis BVBA
Product category
device
Product type
Devices
Recall status
Ongoing — the recall is still in progress
Initiated by
Voluntary: Firm initiated
Distribution
US Nationwide distribution in the state of TX.
Barcode (EAN/GTIN)
05407008320836
Reported risk: Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
What to do

Check whether your product matches this notice (name, batch, dates). If it does, stop using it and follow the instructions in the official notice — refund, return, repair or disposal are set by the issuing authority and the business, and they differ per recall. Official notice from U.S. FDA — Device Enforcement →

Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.