Cardio Vascular-Allura Centron; Model Numbers: 722400; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-09-03
- Country of alert
- Netherlands
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
What the authority reported
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-07-16 · US FDA — Medical Devices
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222,722223 - 2026-07-16 · US FDA — Medical Devices
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232, 722233, 722234, 722235, and 722236, with Soft
All 73 official recall notices naming PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. on this site →
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.