Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in pat

US FDA — Medical Devices · 2024-08-01 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-08-01
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00850001011112
Brand on record
Outset Medical, Inc.

What the authority reported

Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →