NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-09-09
- Country of alert
- United Kingdom
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 05060541640009
- Brand on record
- NOXBOX LTD
What the authority reported
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.