4500MD Triple Quad Mass Spectrometer, Part Number 5032522
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-08-30
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00842365102049
- Brand on record
- AB Sciex, LLC
What the authority reported
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.