4500MD Triple Quad Mass Spectrometer, Part Number 5032522

US FDA — Medical Devices · 2023-08-30 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-08-30
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00842365102049
Brand on record
AB Sciex, LLC

What the authority reported

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.