Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299

US FDA — Medical Devices · 2023-09-07 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-09-07
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00885556773338
Brand on record
Blue Belt Technologies, Inc

What the authority reported

CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters