Official recall record
Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical imaging and non-ima
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-09-06
- Country of alert
- Belgium
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 05400874000710
- Brand on record
- Agfa Healthcare NV
What the authority reported
Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.
What to do