Medfusion Syringe Pump, Model 4000-XXXX-XX: a) 0100-50, b) 0100-249, c) 0101-50, d) 0101-51, e) 0101-78, f) 0101-249, g) 0105-5

US FDA — Medical Devices · 2023-08-16 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-08-16
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10610586035326
Brand on record
Smiths Medical ASD Inc.

What the authority reported

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (including Force Sensor BGND Test, Force Sensor Bridge Test, or Force Sensor Test). However, if the calibration shift is not large enough to trigger a System Failure Alarm, the threshold to detect an occlusion may increase, increasing the time to occlusion detection, or the threshold may decrease, leading to false occlusion alarms. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly. Out of an abundance of caution, we are notifying all customers of this potential issue.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

All 34 official recall notices naming Smiths Medical ASD, Inc. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.