Official recall record
DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-09-02
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 14987666545089
- Brand on record
- Beckman Coulter Inc.
- Product category
- Medical Device
What the authority reported
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-10-23.
Same barcode in other registries
The same EAN appears in 2 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2025-08-04 · US FDA — Medical Devices
DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part - 2026-07-15 · US FDA — Medical Devices
Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, D
Other recalls from this brand
- 2026-09-02 · US FDA — Medical Devices
DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520 - 2026-07-15 · US FDA — Medical Devices
Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Modul - 2026-07-15 · US FDA — Medical Devices
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,
All 32 official recall notices naming Beckman Coulter, Inc. on this site →