Leica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c) 14901850001R0501, d) 1491850UVUL; for

US FDA — Medical Devices · 2024-09-04 · GermanyClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-09-04
Country of alert
Germany
Alert level
Class II
Alert type
recall
Brand on record
LEICA BIOSYSTEMS NUSSLOCH GMBH

What the authority reported

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.