Official recall record
Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2026-08-20
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00763000848156
- Brand on record
- Medtronic MiniMed, Inc.
What the authority reported
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-10-22.
Other recalls from this brand
- 2026-07-13 · US FDA — Medical Devices
MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782; - 2026-07-13 · US FDA — Medical Devices
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712; - 2026-07-13 · US FDA — Medical Devices
MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW