ProductGuru Recalls

Official recall record

Constellation Vision System, REF: 8065751150

US FDA — Medical Devices · 2023-08-18 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-08-18
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00380657511501
Brand on record
Alcon Research, LLC

What the authority reported

Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

All 5 official recall notices naming Alcon Research LLC on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.