Official recall record
DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-09-12
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 15099590732103
- Brand on record
- Beckman Coulter, Inc.
- Product category
- Medical Device
What the authority reported
The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with the results of sample B.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 6 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2024-11-13 · US FDA — Medical Devices
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137 - 2025-04-23 · US FDA — Medical Devices
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software version 1.20 - 2025-09-19 · US FDA — Medical Devices
DxI 9000 Access Immunoassay Analyzer, Part Number C11137 - 2024-03-05 · US FDA — Medical Devices
DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, - 2024-01-03 · US FDA — Medical Devices
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument
Other recalls from this brand
- 2026-09-21 · US FDA — Medical Devices
ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima X - 2026-09-21 · US FDA — Medical Devices
ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima X - 2026-08-12 · US FDA — Medical Devices
DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
All 39 official recall notices naming Beckman Coulter, Inc. on this site →