ProductGuru Recalls

Official recall record

IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set

US FDA — Medical Devices · 2025-10-02 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-10-02
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
10887709007823
Brand on record
ICU Medical, Inc.
Product category
Medical Device

What the authority reported

IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.