ProductGuru Recalls

Official recall record

IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ20

US FDA — Medical Devices · 2026-08-12 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-08-12
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10837461001256
Brand on record
Beckman Coulter, Inc.
Product category
Medical Device

What the authority reported

The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-11-12.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

All 39 official recall notices naming Beckman Coulter, Inc. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.