Official recall record
TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-09-16
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 07613327548082
- Brand on record
- Howmedica Osteonics Corp.
What the authority reported
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-11-12.
Record last checked against the registry on 2026-11-12.
Other recalls from this brand
- 2026-09-16 · US FDA — Medical Devices
TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519. - 2026-09-16 · US FDA — Medical Devices
TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507. - 2026-09-16 · US FDA — Medical Devices
Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.
All 20 official recall notices naming Howmedica Osteonics Corp. on this site →
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.