Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog

US FDA — Medical Devices · 2023-09-15 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-09-15
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
00815440022075
Brand on record
Asensus Surgical Inc

What the authority reported

Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.