Techno-Path Multichem AE Unassayed Level 3, REF:AE623X

US FDA — Medical Devices · 2026-08-14 · IrelandClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-08-14
Country of alert
Ireland
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
05391523442824
Brand on record
TECHNO-PATH MANUFACTURING LTD.

What the authority reported

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-11-13.

More recalls: US FDA — Medical Devices · Ireland · all filters

Other recalls from this brand

All 5 official recall notices naming TECHNO-PATH MANUFACTURING LTD. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.