MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Pr

US FDA — Medical Devices · 2025-08-22 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-08-22
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00840682125499
Brand on record
GE Medical Systems, LLC

What the authority reported

If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.