MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Pr
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-08-22
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00840682125499
- Brand on record
- GE Medical Systems, LLC
What the authority reported
If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-06-05 · US FDA — Medical Devices
Optima XR200amx, Mobile Digital-Ready Radiographic System - 2026-06-05 · US FDA — Medical Devices
Optima XR240amx, Mobile Digital Radiographic System - 2026-06-05 · US FDA — Medical Devices
Optima XR220amx, Mobile Digital-Ready Radiographic System
All 62 official recall notices naming GE Medical Systems, LLC on this site →
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.