Official recall record
VariSoft infusion set, Model Number 1002828
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-10-11
- Country of alert
- Denmark
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 05705244018730
- Brand on record
- Unomedical A/S
What the authority reported
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.