Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091

US FDA — Medical Devices · 2024-09-12 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-09-12
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00884999048614
Brand on record
Abbott Molecular, Inc.

What the authority reported

Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.

What to do