ProductGuru Recalls

Official recall record

VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartri

US FDA — Medical Devices · 2023-09-12 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-09-12
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10758750009510
Brand on record
Ortho-Clinical Diagnostics, INc.
Product category
Medical Device

What the authority reported

Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.