ProductGuru Recalls

Official recall record

17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-O

US FDA — Medical Devices · 2024-09-13 · United StatesClass III
Issuing authority
US FDA — Medical Devices
Alert date
2024-09-13
Country of alert
United States
Alert level
Class III
Alert type
recall
Barcode (EAN/GTIN)
04048474012927
Brand on record
DRG International, Inc.

What the authority reported

The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.