ProductGuru Recalls

Official recall record

AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.

US FDA — Medical Devices · 2024-09-27 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-09-27
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10711234530023
Brand on record
Accriva Diagnostics, Inc.
Product category
Medical Device

What the authority reported

Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand