ProductGuru Recalls

Official recall record

Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomog

US FDA — Medical Devices · 2024-09-17 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-09-17
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
04987670103810
Brand on record
Canon Medical System, USA, INC.

What the authority reported

When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.