Official recall record
Millenium Myosight; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2025-09-12
- Country of alert
- Israel
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
What the authority reported
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2025-12-24 · US FDA — Medical Devices
GE Healthcare Omni Legend, sold under the following names: OMNI Legend 16, Omni Legend 21, OMNI Legend 32, - 2025-12-10 · US FDA — Medical Devices
SMV DSI, System, Tomography, Computed, Emission - 2025-09-12 · US FDA — Medical Devices
Starcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?