IRON assay, Reference Numbers 6K95-41 and 6K95-30
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-09-29
- Country of alert
- Italy
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00380740010928
- Brand on record
- Sentinel CH SpA
What the authority reported
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.