ProductGuru Recalls

Official recall record

Rhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea).

US FDA — Medical Devices · 2023-09-25 · JapanClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-09-25
Country of alert
Japan
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
04953170307188
Brand on record
Aizu Olympus Co., Ltd.

What the authority reported

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Japan · all filters

Other recalls from this brand

All 18 official recall notices naming Aizu Olympus Co., Ltd. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.