Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL,

US FDA — Medical Devices · 2023-11-01 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-11-01
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
04046964837807
Brand on record
B. Braun Medical, Inc.

What the authority reported

In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

All 46 official recall notices naming B. Braun Medical, Inc. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.