Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL,
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-11-01
- Country of alert
- United States
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 04046964837807
- Brand on record
- B. Braun Medical, Inc.
What the authority reported
In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2024-06-17 · US FDA — Medical Devices
OUTLOOK PUMP SET 3 CARESITE LADS 133 IN.-Used with an electrically-powered infusion pump for intravenous (IV) - 2024-06-17 · US FDA — Medical Devices
Infusomat SPACE PUMP IV SET, 127 IN.-Used with an electrically-powered infusion pump for intravenous (IV) adm - 2024-06-17 · US FDA — Medical Devices
Infusomat 60DROP METRISET PUMP SET, 3 SAFELINE-Used with an electrically-powered infusion pump for intravenou
All 46 official recall notices naming B. Braun Medical, Inc. on this site →