ProductGuru Recalls

Official recall record

DigitalDiagnost C90 (1) High Performance, (2) Flex/Value/Chest/ER; Model Number: (1) 712034, (2) 712035;

US FDA — Medical Devices · 2025-10-01 · GermanyClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-10-01
Country of alert
Germany
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00884838090699
Brand on record
Philips Medical Systems DMC GmbH

What the authority reported

Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Germany · all filters

Same barcode in other registries

The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand