Official recall record
Defibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-10-14
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00815098020447
- Brand on record
- Defibtech, LLC
What the authority reported
Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life.there is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2025-03-18 · US FDA — Medical Devices
Automated External Defibrillators, Model DDU-2450. Item number: DCF-E2460DE (German language) - 2025-03-18 · US FDA — Medical Devices
Automated External Defibrillators, Model DDU-2200, IItem numbers: DCF-E2210-D3/1(German/English dual language) - 2025-03-18 · US FDA — Medical Devices
Semi-Automated External Defibrillators; Models: DDU-100 and DDU-100E. Item numbers: DCF-E110SG-DE/1 (German la