ProductGuru Recalls

Official recall record

Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 39

US FDA — Medical Devices · 2023-10-25 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-10-25
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00840277402172
Brand on record
Riverpoint Medical, LLC

What the authority reported

Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.