Official recall record
smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-10-28
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 35021239002579
- Brand on record
- Smiths Medical ASD, Inc.
What the authority reported
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2025-02-14 · US FDA — Medical Devices
CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214 - 2025-02-14 · US FDA — Medical Devices
CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW1100