Official recall record
Ivenix Infusion System (IIS), Large Volume Pump LVP-0004
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-11-29
- Country of alert
- United States
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00811505030320
- Brand on record
- Fresenius Kabi USA, LLC
What the authority reported
The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the issue may lead to serious harm or death.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 6 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2025-11-03 · US FDA — Medical Devices
Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004. - 2024-12-06 · US FDA — Medical Devices
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004 - 2023-03-10 · US FDA — Medical Devices
Ivenix Infusion System (IIS), Large Volume Pump, Model No. LVP-0004 - 2025-09-03 · US FDA — Medical Devices
Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004. - 2026-02-04 · US FDA — Medical Devices
The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.
Other recalls from this brand
- 2026-05-06 · US FDA — Medical Devices
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.