Ivenix Infusion System (IIS), Large Volume Pump LVP-0004

US FDA — Medical Devices · 2023-11-29 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-11-29
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
00811505030320
Brand on record
Fresenius Kabi USA, LLC

What the authority reported

The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the issue may lead to serious harm or death.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.