Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

US FDA — Medical Devices · 2024-10-09 · JapanClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-10-09
Country of alert
Japan
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
04540217420582
Brand on record
Shimadzu Corp. Analytical & Measuring Instruments Div.

What the authority reported

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.