Ziehm Solo FD. Interventional fluoroscopic x-ray system

US FDA — Medical Devices · 2024-08-08 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-08-08
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Orthoscan, Inc.

What the authority reported

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.