Official recall record
MICRO-X Rover Mobile X-ray System, # MXU-RV35
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-10-22
- Country of alert
- Australia
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 9357123000037
- Brand on record
- Micro-X Ltd.
- Product category
- Medical Device
What the authority reported
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.