Atellica CI Analyzer. Catalog Numbers: 10947347.

US FDA — Medical Devices · 2024-11-04 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-11-04
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
630414229560
Brand on record
Siemens Healthcare Diagnostics, Inc.

What the authority reported

Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.