ProductGuru Recalls

Official recall record

Alinity ci-series System Control Module, REF: 03R70-01

US FDA — Medical Devices · 2023-05-30 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-05-30
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00380740137410
Brand on record
Abbott Laboratories

What the authority reported

There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3) Customer requests mechanism to prevent user error when creating new lots of value assigned Calibrators; 4) Error in processing tests after CCCintController encountered an error with the assay information downloaded from SCC; 5) The assay editor does not correctly check sample dilution total volume limits; 6) User-Applied Labels not adhering to Reagent Bottles; 7) ICT Module did not expire after warranty was exceeded. These performance issues could lead to erroneous results for multiple analytes.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 6 official recall notices naming Abbott Laboratories on this site →