Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000

US FDA — Medical Devices · 2023-11-15 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-11-15
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Landauer

What the authority reported

MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.