ProductGuru Recalls

Official recall record

StatStrip Glucose Ketone (mmol/L) Hospital Meter System. Model/Catalog Number: 63910. Prescription use blood glucose ketone meter for near-p

US FDA — Medical Devices · 2024-11-20 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Alert date
2024-11-20
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
1038548063910
Brand on record
Nova Biomedical Corporation

What the authority reported

A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.