ProductGuru Recalls

Official recall record

Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2

US FDA — Medical Devices · 2023-12-05 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-12-05
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Ambu Inc.

What the authority reported

Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.