Official recall record
GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-11-25
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00840682115278
- Brand on record
- GE Medical Systems, LLC
What the authority reported
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2026-06-05 · US FDA — Medical Devices
Optima XR200amx, Mobile Digital-Ready Radiographic System
Other recalls from this brand
- 2026-06-05 · US FDA — Medical Devices
Optima XR200amx, Mobile Digital-Ready Radiographic System - 2026-06-05 · US FDA — Medical Devices
Optima XR240amx, Mobile Digital Radiographic System - 2026-06-05 · US FDA — Medical Devices
Optima XR220amx, Mobile Digital-Ready Radiographic System