Official recall record
TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-12-18
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- St. Jude Medical, Atrial Fibrillation Division, Inc.
What the authority reported
When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.