ProductGuru Recalls

Official recall record

GE Voluson IC9-RS intracavitary probes, Model Number J48691PJ, Diagnostic ultrasound imaging and fluid flow analysis

US FDA — Medical Devices · 2023-12-22 · AustriaClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-12-22
Country of alert
Austria
Alert level
Class II
Alert type
recall
Brand on record
GE HEALTHCARE AUSTRIA GMBH & CO

What the authority reported

GE HealthCare has become aware of an issue with certain IC9-RS ultrasound probes (see Affected Products List) that can result in a double image artifact creating a ghost image with realistic features. An unrecognized artifact may lead to misdiagnosis.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Austria · all filters