ARTIS One Angiographic X-Ray System

US FDA — Medical Devices · 2025-01-02 · GermanyClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-01-02
Country of alert
Germany
Alert level
Class II
Alert type
recall
Brand on record
SIEMENS MEDICAL SOLUTIONS USA, INC

What the authority reported

A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.