ARTIS One Angiographic X-Ray System
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-01-02
- Country of alert
- Germany
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- SIEMENS MEDICAL SOLUTIONS USA, INC
What the authority reported
A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.