BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 4535602

US FDA — Medical Devices · 2023-12-15 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-12-15
Country of alert
United States
Alert level
Class I
Alert type
recall
Brand on record
Philips North America

What the authority reported

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.