Official recall record
BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyx
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-11-20
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 10885403512667
- Brand on record
- CareFusion 303, Inc.
- Product category
- Medical Device
What the authority reported
Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 6 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2024-02-01 · US FDA — Medical Devices
BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352; - 2025-01-08 · US FDA — Medical Devices
(1) Pyxis MedStation ES, Catalog No. 323; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) Pyxis MedStation ES - 2024-05-02 · US FDA — Medical Devices
BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES - 2024-11-14 · US FDA — Medical Devices
BD Pyxis MedStation ES, REF: 323 - 2023-04-05 · US FDA — Medical Devices
BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352
Other recalls from this brand
All 11 official recall notices naming CareFusion 303, Inc. on this site →