ProductGuru Recalls

Official recall record

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

US FDA — Medical Devices · 2025-11-24 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-11-24
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00887709127227
Brand on record
ICU Medical, Inc.

What the authority reported

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 16 official recall notices naming ICU Medical, Inc. on this site →