BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

US FDA — Medical Devices · 2025-12-05 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-12-05
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00886158037576
Brand on record
Beaver-Visitec International, Inc.

What the authority reported

Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).