Official recall record
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-12-05
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00886158037576
- Brand on record
- Beaver-Visitec International, Inc.
What the authority reported
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).